BT100-1F Low Speed Peristaltic Dispensing Pump
The BT100-1F peristaltic dispensing pump is mainly designed for laboratory use.
This dispensing pump delivers flow rates up to 500 ml/min and dispensing volume ranges from 0.01 ml to 9900 ml.
Combining a 128×32 LCD with a membrane keypad and a rotary encoded switch makes the operation easy and prompt.
The speed can be adjusted manually or automatically through an external control interface.
Specifications
- Speed: 0.1 to 100 rpm, reversible
- Flow rates: 0.2μL to 500 ml/min
- Dispensing volume: 0.01 ml to 9.99 litres
- Copy number: 0 to 9999, "0" means unlimited cycle
- Pause time: 0.1 sec. to 99.9 min, resolution is 0.1 sec.
- Back suction time: 0 to 99.9 sec., resolution is 0.1 sec.
- External control: Start/stop, direction and speed control (0-5V, 0-10V, 4-20 mA and 0-10 kHz control modules are optional)
- Speed signal output: 0-10 kHz corresponding to 0.1-100 rpm
- Start/stop signal output: OC gate signal output
- Direction signal output: OC gate signal output
- Communication interface: RS485
Functions and features
- Micro-flow rates transferring: The smallest flow rate can reach 0.2 μl/min
- Back suction function: In dispensing mode, the pump reverses after finishing one filling operation to prevent the liquid from dripping
- Operating mode: Membrane keypad and rotary encoded switch
- Speed control: The speed can be adjusted manually or automatically through external control interface
- Display: LCD displays current running status
- Prime key: For fast filling and emptying
- Memory function: Storing the running parameters automatically
- Calibration function: The flow rates and the dispensing volume can be calibrated to increase the accuracy
- Communication function: Computer control available through RS485 communication interface
Operating Manual
- Supplied with English operating manual
Parameters
| Drive |
Pump head |
Tubing size # |
Flow rate (ml/min/channel) at 0.1 - 100 rpm (measured with water) |
Weight (kg) |
| BT100-1F |
YZ1515X or YZII15
with 3 rollers |
13 |
0.006 - 6.3 |
4.2 |
| 14 |
0.025 - 25 |
| 19 |
0.042 - 41.66 |
| 16 |
0.077 - 77.7 |
| 25 |
0.16 - 160 |
| 17 |
0.26 - 266 |
| 18 |
0.36 - 366 |
YZ2515X or YZII25
with 3 rollers |
15 |
0.16 - 160 |
| 24 |
0.26 - 266 |
YZII25
with 3 rollers |
35 |
0.36 - 360 |
| 36 |
0.5 - 500 |
DG-1, 2 or 4
with 6 rollers and 1, 2 or 4 channels |
Tubing with stops
WT: 0.8-1.0 mm,
ID: ≤ 3.17 mm |
Max.48 |
4 - 4.3 |
DG-1, 2 or 4
with 10 rollers and 1, 2 or 4 channels |
Max. 32 |
| DMD15-13-B |
13 |
0.013 - 13.3 |
4.2 |
| 14 |
0.05 - 50 |
| 19 |
0.103 - 103.3 |
| 16 |
0.16 - 160 |
| 25 |
0.35 - 350 |
DG15-24
with 2 channels and 4 rollers |
16 |
0.082 - 81.7 |
4.6 |
| 25 |
0.192 - 191.7 |
| 17 |
0.3 - 300.0 |
DG15-28
with 2 channels and 8 rollers |
13, 14
WT: 0.8 - 1.0 (mm)
ID: ≤ 3.17 mm |
Max. 75 (pr channel) |
4.5 |
*WT: Wall thickness, ID: Inner diameter
Note: If you choose a LP-YZ1515x with 6 rollers the flow rate will be 2/3 of the flow rate LP-YZ1515x with 3 rollers.
VIDEOS
Dispensing Demo
Dispensing Calibration
Flow Rate Calibration
Back Suction Demo