Syringe Pump LP-LSP01-2A

Syringe Pump LP-LSP01-2A

EUR 2,064.90

(excl. VAT)

How to install the syringe and to make the settings on the syringe pump LP-LSP

 

Read more
Model/Product no.: LP-LSP01-2A
Weight: 3.6 kg
Stock status: approx. 3 weeks delivery
pcs.

Single channel syringe pump with infusion mode

This pump is a single channel syringe pump which only has infusion mode.
The acceptable syringe specification ranges from 10 μl to 60 ml.
Suitable for high accuracy and small flow rate liquid transferring.

Specifications

  • Max. no. of syringes: 1
  • Pusher advance per micro-step: 0.03125 µm
  • Syringe size:  10 µl - 60 ml
  • Flow rates: 0.0002 μl - 8.6699 ml/min
  • Max. linear rate: 13 mm/min
  • Min. linear rate: 1 µm/min
  • Linear speed resolution: 1 µm/min
  • Max. step rate: 6933 (1/16 step)/sec
  • Min. step rate: 16 (1/16 step)/30 sec
  • Working mode: Infusion
  • Linear force:  > 9kg
  • Stroke of pump: 140 mm
  • Accuracy:  ≤ ± 0.5% error in the condition of > 30% of max. pump stroke
  • Operating mode: Rotary encoded switch and membrane keypad
  • Display: 128 x 64 graphic LCD
  • Power: AC 90V-260V/15W
  • Operating condition: Temperature 5-40°C,  Relative humidity < 80 %
  • Dimensions:  280×210×140 (mm) 280×250×140 (mm)
  • Weight:  3.6 kg

Functions and features

  • Syringe selection: The syringe can be selected in the manufacturer table which includes manufacturer, material and size or the inner diameter of the syringe barrel can be input directly
  • Easy to operate: Combining big screen LCD display with a rotary encoded switch and membrane keypad makes the operation simple and prompt
  • Working mode: Infusion
  • Memory function: 
    • The parameters are saved in EEPROM. The parameters do not need to be reset when power is restored after an interruption
    • In flow rate mode, the pump continues running or stops according to the setting parameters when power is restored after an interruption
  • Protection function: The pump will stall and an alarm will sound when the drive structure of the pump is blocked
  • Communication function: Computer control available through RS485 communication interface
  • External control function: Input/output control
  • Calibration function: Acquire accurate volume through calibration
  • Syringe protection: Adjust syringe rest to prevent syringe from damage

Parameters

Type Syringe     Syringe ID (mm) Flow Rate (μl/min-ml/min)
LSP01-2A 10μl 0.50 0.0002-0.0026
25μl 0.80 0.0005-0.0065
50μl 1.10 0.0095-0.0124
100μl 1.60 0.002-0.026
250μl 2.30 0.0042-0.054
500μl 3.25 0.0083-0.1078
1ml 4.72 0.0175-0.2275
2ml 9.00 0.0636-0.827
5ml 13.10 0.1348-1.7522
10ml 16.60  0.2164-1.8135
20ml 19.00  0.2835-3.6859
30ml 23.00  0.4155-5.4012
60ml 29.14  0.6669-8.6699

 

How to make the basic settings on the syringe pump LP-LSP

Calibration of the flow rate on the syringe pump LP-LSP

 

Weight 3.6 kg
CE marking Yes
Operating Temperature 5-40°C
Relative humidity <80%
Size (Length) 280 mm
Size (Height) 140 mm
Size (Width) 210 mm