Thank you

The following product(s) are added to the basket

Buy More
     

LP-LSP01-1BH - Syringe pump

Part Number LP-LSP01-1BH
LP-LSP01-1BH - Syringe pump

Single channel high pressure syrings pump. Flow rate: 0.09 µl-124.36 ml/min. Linear force: 45 kg

Description:

The LP-LSP01-1BH pump is a single channel high pressure syringe pump.

The acceptable syringes are 2.5 ml - 100 ml stainless steel syringes.

Because of the extra linear force this pump is suitable for transferring viscous fluids or delivering fluids to reactors in chemical applications.

Specifications
Syringe size: 2.5 ml - 100 ml
Linear force: >450 N (45 kg)
Advance per micro-step:  0.156 μm (100 ml syringe 1/16 step)
Infusion volume per micro-step:  0.149 μl (100 ml syringe 1/16 step)
Max. step rate: 13867 (1/16 step)/sec
Min. step rate: 16 (1/16 step)/30 sec
Max. linear rate: 130 mm/min
Min. linear rate: 5 μm/min
Flow rates: 0.09 μl/min-124.36 ml/min
Accuracy: ≤±0.5% error in the condition of ≥30% of max. infusion distance
Setting mode: Rotary coded switch and membrane keypad
Display: 128×64 graphic LCD
Power: AC 100-240V
Operating condition: Temperature 5-40 °C, Relative humidity <80%
Dimensions: 280×250×140 (mm)
Weight: 6.3 kg


Functions and features

Syringe selection:
The syringe can be selected in the manufacturer table which includes manufacturer, material and size or the inner diameter of the syringe barrel can be input directly.

Easy to operate:
Combining big screen LCD display with a rotary encoded switch and membrane keypad makes the operation simple and prompt.

Work mode:
Infusion, withdrawal, infusion/withdrawal, withdrawal/infusion and continuous.

Memory function:

  1. The parameters are saved in EEPROM.The parameters don't need to be reset when power returns after an interruption.
  2. In flow rate mode, the pump continues running or stops according to the setting parameters when power returns after an interruption.

Protection function:
The pump will stall and an alarm will sound when the drive structure of the pump is blocked.

Communication function:
Computer control available through RS485 communication interface.

External control function:
Input/output control

Calibration function:
Acquire accurate volume through calibration.

Syringe protection:
Adjust the syringe rest to prevent damage to the syringe.

Parameters
Syringe pump Syringe Inner diameter (mm) Ref. flow rate (*/hr) Outlet pressure Linear rate (/min) Weight (kg)
LSP01-1BH 2.5 ml 4.79 5.406 µl-140.5 ml >19.48 MPa 5 µm-130 mm  6.3    
8 ml 9.53 21.399 µl-556.3 ml >5.05 MPa
20 ml 19.13 86.226 µl-2241.8 ml >1.25 MPa
50 ml 28.6 192.727 µl-5010.9 ml >0.56 MPa
100 ml 34.9 286.986 µl-7461.6 ml >0.37 MPa

EUR 4,277.81




Copyright 2011 - Drifton A/S - CVR / VAT no.: 53284310